The new NMi orthovolt x-rays absorbed dose to water primary standard based on water calorimetry

A new water calorimeter for orthovolt x-rays has been developed at NMi. The purpose of this calorimeter is to provide calibrations in terms of absorbed dose to water. Four internationally accepted CCRI qualities (100-250 kV) have been characterized. Correction factors have been determined with measu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics in medicine & biology 2008-07, Vol.53 (13), p.3531-3542
Hauptverfasser: de Prez, L A, de Pooter, J A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new water calorimeter for orthovolt x-rays has been developed at NMi. The purpose of this calorimeter is to provide calibrations in terms of absorbed dose to water. Four internationally accepted CCRI qualities (100-250 kV) have been characterized. Correction factors have been determined with measurements, Monte Carlo calculations and heat transport models. The results of the calculations have been validated against measurements. The absorbed dose to water calibration coefficient, N(Dw,WCM), with the new calorimeter has been compared to the N(Dw,NCS) based on an air-kerma calibration coefficient, using the current NCS-10 dosimetry protocol. A good agreement is found for all beam qualities where the total uncertainty (1 SD) has decreased from 2.5% for N(Dw,NCS) to better than 1.5% for N(Dw,WCM), using the new water calorimeter.
ISSN:0031-9155
1361-6560
DOI:10.1088/0031-9155/53/13/009