Repeatability and agreement of two A-scan ultrasonic biometers and IOLMaster in non-orthokeratology subjects and post-orthokeratology children
Purpose: Our aim was to determine the repeatability of measurements of axial length (AL) and anterior chamber depth (ACD) made with two ultrasonic biometers and the IOLMaster in a group of non-orthokeratology (ortho-k) adult subjects and to investigate the agreement among instruments in children un...
Gespeichert in:
Veröffentlicht in: | Clinical and experimental optometry 2006-05, Vol.89 (3), p.160-168 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose: Our aim was to determine the repeatability of measurements of axial length (AL) and anterior chamber depth (ACD) made with two ultrasonic biometers and the IOLMaster in a group of non-orthokeratology (ortho-k) adult subjects and to investigate the agreement among instruments in children undergoing ortho-k therapy and in children wearing spectacles.
Methods: To determine repeatability, AL and ACD were measured twice in 22 non-ortho-k young adults using two A-scan ultrasonic biometers (A-5500 and A-2500) and the IOLMaster. To determine agreement, AL and ACD were measured with the same instruments in 30 children undergoing ortho-k therapy and 30 spectacle-wearing children.
Results: In the adult subjects, there were no significant differences in ACD and AL measurements obtained from the three instruments (repeated measures ANOVAs, p > 0.05). There was also no significant between-measurement difference for each instrument. The between-measurement agreement was better for the IOLMaster (95% limits of agreement (LA): -0.04 and +0.05-mm for both AL and ACD) than for the two A-scan ultrasonic biometers (95% LA: -0.12 and +0.11-mm for AL; -0.22 and +0.27-mm for ACD). Among the children, AL measurements with all three instruments were not significantly different from each other for both the children undergoing ortho-k therapy and those wearing spectacles (repeated measures ANOVAs, p > 0.05). The 95% LA of differences obtained from any two instruments were also comparable for both groups of subjects (within -0.20-mm and +0.20-mm). ACD measurements of the children were significantly different among the three instruments (repeated measures ANOVAs, p 0.017).
Conclusions: The repeatability of AL and ACD measurements with the IOLMaster was very good, and was better than with the A-scan ultrasonic biometers. The agreements in AL measurements between A-scan ultrasonic biometers and IOLMaster were comparable in both the ortho-k and the spectacle-wearing subjects, and were comparable to the repeatability of the A-scan ultrasonic biometers. ACD measurements between A-scan ultrasonic biometry and the IOLMaster were not comparable. AL measurements with the IOLMaster can replace the measurements from the two A-scan ultrasonic biometers used, however, the reverse is not true. AL and ACD measurements with all three instruments wer |
---|---|