Determination of valdecoxib in serum using a HPLC-diode array detector and its application in a pharmacokinetic study

A simple, sensitive, isocratic and reproducible reversed phase HPLC method for the determination of valdecoxib, a novel specific COX-2 inhibitor in human serum was developed using a diode array detector and celecoxib as internal standard. The system consisted of a C18 column and a detector set at 24...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pharmazie 2006-03, Vol.61 (3), p.245-246
Hauptverfasser: Keshetty, S., Venisetty, R. K., Molmoori, V., Ciddi, V.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A simple, sensitive, isocratic and reproducible reversed phase HPLC method for the determination of valdecoxib, a novel specific COX-2 inhibitor in human serum was developed using a diode array detector and celecoxib as internal standard. The system consisted of a C18 column and a detector set at 240 nm. The mobile phase was a mixture of acetonitrile : water acidified to pH 3.2 with orthophosphoric acid (OPA) (60 : 40) pumped at room temperature and a flow rate of 1 ml/min. The mean absolute recovery value was about 90%, while the intra (n = 5) and inter (n = 5) assay variations were 0.999. The limit of detection was ≤10 ng/ml. The method was used to study the pharmacokinetics of valdecoxib after a single dose oral administration to human volunteers.
ISSN:0031-7144