Noise and spatial resolution of a real-time electrical impedance tomograph
The Sheffield real-time electrical impedance tomograph produces 25 images per second, using 16 electrodes with adjacent-pair current drive and non-iterative image reconstruction. We describe the data acquisition timing of this instrument and present quantitative measurements of its signal-to-noise r...
Gespeichert in:
Veröffentlicht in: | Clinical physics and physiological measurement 1992-12, Vol.13 (A), p.125-130 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Sheffield real-time electrical impedance tomograph produces 25 images per second, using 16 electrodes with adjacent-pair current drive and non-iterative image reconstruction. We describe the data acquisition timing of this instrument and present quantitative measurements of its signal-to-noise ratio and spatial resolution. |
---|---|
ISSN: | 0143-0815 |