A single large dose of megavoltage x-rays delivered through a grid is currently being utilized by some centres for palliative radiotherapy treatments of large tumours. In this note, we investigate the dosimetry of grid therapy using two-dimensional film dosimetry and three-dimensional gel dosimetry....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics in medicine & biology 2004-10, Vol.49 (19), p.N317-N323
Hauptverfasser: Trapp, J V, Warrington, A P, Partridge, M, Philps, A, Glees, J, Tait, D, Ahmed, R, Leach, M O, Webb, S
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A single large dose of megavoltage x-rays delivered through a grid is currently being utilized by some centres for palliative radiotherapy treatments of large tumours. In this note, we investigate the dosimetry of grid therapy using two-dimensional film dosimetry and three-dimensional gel dosimetry. It is shown that the radiation dose is attenuated more rapidly with depth in a grid field than an open field, and that even shielded regions receive approximately 25% of the dose to the unshielded areas.
ISSN:0031-9155