A Simple Method to Assess Accuracy of Deep Brain Stimulation Electrode Placement: Pre-Operative Stereotactic CT + Postoperative MR Image Fusion
Objective: To describe a method for the measurement of the accuracy of deep brain stimulation (DBS) electrode placement with the use of image fusion technologies. Patients and Methods: Ten consecutive patients suffering from movement disorders underwent DBS electrode placement. Postoperative MR imag...
Gespeichert in:
Veröffentlicht in: | Stereotactic and functional neurosurgery 2004-01, Vol.82 (1), p.14-19 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective: To describe a method for the measurement of the accuracy of deep brain stimulation (DBS) electrode placement with the use of image fusion technologies. Patients and Methods: Ten consecutive patients suffering from movement disorders underwent DBS electrode placement. Postoperative MR images were fused with the pre-operative stereotactic CT. The placement error in the anteroposterior, lateral and vertical planes was calculated. Results: The anteroposterior mean error ± SD was 0.61 ± 0.22 mm (range 0.2–0.9 mm). The lateral mean error ± SD was 0.65 ± 0.27 mm (range 0.2–2.2 mm). The vertical mean error ± SD was 0.82 ± 0.31 mm (range 0.3–1.6 mm). Conclusions: This technique provides a simple and precise method for the evaluation of the accuracy of DBS electrode placement. |
---|---|
ISSN: | 1011-6125 1423-0372 |
DOI: | 10.1159/000076655 |