Inaccuracy of lipid measurements with the portable Cholestech L.D.X analyzer in patients with hypercholesterolemia
Although total cholesterol concentrations measured by portable lipid analyzers have acceptable bias and precision in young and middle-aged adults, clinically relevant differences in HDL-cholesterol (HDL-C) and triglyceride values have been described. Furthermore, the accuracy of portable lipid analy...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2002-02, Vol.48 (2), p.284-290 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Although total cholesterol concentrations measured by portable lipid analyzers have acceptable bias and precision in young and middle-aged adults, clinically relevant differences in HDL-cholesterol (HDL-C) and triglyceride values have been described. Furthermore, the accuracy of portable lipid analyzers in older hyperlipidemic individuals, who have a high incidence of coronary heart disease, has not been validated. This study determined the biases and variability in portable lipid measurements in older patients with hypercholesterolemia and related them to National Cholesterol Education Program Adult Treatment Panel III guidelines.
Participants were > or =70 years of age with fasting serum LDL-cholesterol (LDL-C) concentrations > 1.40 g/L. Fasting fingerstick samples were analyzed on a Cholestech L.D.X desktop analyzer. Antecubital venous samples were analyzed in a proficiency-certified clinical laboratory.
Portable measurements systematically overestimated triglycerides (0.296 g/L; P |
---|---|
ISSN: | 0009-9147 1530-8561 |