Detecting electron beam energy shifts with a commercially available energy monitor

Routine electron beam quality assurance requires an accurate, yet practical, method of energy characterization. Subtle shifts in beam energy may be produced by the linac bending magnet assembly, and the sensitivity of a commercially available electron beam energy-monitoring device for monitoring the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2000-07, Vol.27 (7), p.1661-1663
Hauptverfasser: Evans, Michael D. C., Moftah, Belal A., Olivares, Marina, Podgorsak, Ervin B.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Routine electron beam quality assurance requires an accurate, yet practical, method of energy characterization. Subtle shifts in beam energy may be produced by the linac bending magnet assembly, and the sensitivity of a commercially available electron beam energy-monitoring device for monitoring these small energy drifts has been evaluated. The device shows an 11% change in signal for a 2 mm change in the I 50 energy parameter for low energy electron beams (in the vicinity of 6 MeV) and a 2.5% change in signal for a 2 mm change in the I 50 energy parameter for high energy electron beams (in the vicinity of 22 MeV). Thus the device is capable of detecting small energy shifts resulting from bending magnet drift for all clinically relevant electron beams.