Leak detector in medical gas delivery systems

We describe an original device developed by the biomedical technicians of the University Hospital of Nancy for checking tightness and measuring the capacity of medical gas pipeline network. A leak can be assessed either as a loss of pressure or as a loss of volume per unit of time (cmH2O.min-1 and m...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annales françaises d'anesthésie et de réanimation 1999-11, Vol.18 (9), p.1009
Hauptverfasser: Kurtz, M, Labrude, M, Noirez, V
Format: Artikel
Sprache:fre
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We describe an original device developed by the biomedical technicians of the University Hospital of Nancy for checking tightness and measuring the capacity of medical gas pipeline network. A leak can be assessed either as a loss of pressure or as a loss of volume per unit of time (cmH2O.min-1 and mL.min-1 respectively). To assess the latter, the capacity of the explored pipeline must have been measured when it still was gas-tight. With the device, a leak can be assessed within a few minutes whereas the official procedure requires a delay between 1 and 24 hours. The device still has to undergo bench evaluation and validation.