Quantitative optical densitometry with scanning-laser film digitizers
A new process for eliminating two types of artifacts inherent in commercially available transmission scanning-laser film digitizers is presented. The first kind of artifact results in nonreproducible interference-pattern fluctuations as large as 7%. The second kind results in spreading of transmitte...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 1999-08, Vol.26 (8), p.1721-1731 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new process for eliminating two types of artifacts inherent in commercially available transmission scanning-laser film digitizers is presented. The first kind of artifact results in nonreproducible interference-pattern fluctuations as large as 7%. The second kind results in spreading of transmitted light from low-to-high optical density (OD) in regions with rapidly varying ODs, producing errors as large as 50%. These OD artifacts cause the loss of precision for films with low-OD regions (first type) and the loss of accuracy for films with regions of high-OD near high-OD gradients (second type). Test radiochromic films, produced by uniform exposure to a 6 MV photon beam and a high dose rate
192
Ir
brachytherapy source, along with test radiographic films were used to characterize the artifacts of a commercially available scanning-laser film digitizer. The interference-pattern artifact was eliminated by digitizing the films on a masked diffusing ground-glass scanning bed. The light-transmission artifact was eliminated through discrete-fast-Fourier-transform (DFFT) deconvolution of transmission profiles with measured digitizer line-spread functions. Obtaining precise OD distributions after the DFFT deconvolution required prior removal of the interference-pattern artifact and application of a low-pass Wiener noise filter. Light-transmission artifacts are particularly significant for applications requiring measurement of high-gradient OD distributions, such as brachytherapy or conformal photon-beam film dosimetry and quantitation of two-dimensional electrophoresis gels. Errors as large as 15%–35% occur in OD distributions representative of these applications. The data collection and correction process developed in this study successfully removes these artifacts. |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1118/1.598664 |