Localization of a ventricular tachycardia-focus with multichannel magnetocardiography and three-dimensional current density reconstruction
The objective of this case report is to determine the accurate localization of a malignant ventricular tachycardia (VT)focus by combining multichannel magnetocardiographic (MCG) information with morphologic data. The localization was obtained by calculating the three-dimensional current density dist...
Gespeichert in:
Veröffentlicht in: | Journal of medical engineering & technology 1999, Vol.23 (3), p.108-115 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The objective of this case report is to determine the accurate localization of a malignant ventricular tachycardia (VT)focus by combining multichannel magnetocardiographic (MCG) information with morphologic data. The localization was obtained by calculating the three-dimensional current density distribution (3D-CDD) on the left ventricular surface. To estimate the accuracy of this localization technique, examinations of a healthy volunteer were additionally performed. The MCG-signals wererecorded in a magneticallyshielded roombya 49-channel magnetogradiometer. The corresponding morphologic information was recorded by magnetic resonance tomography(MRT). Thecoordinatesystems werematched with thehelp of markers. The3D-CDDwas calculated bythePhilips CURRY software package. The origin of a malignant VT determined by X-ray images of the ablation catheter position during the electrophysiological examination (EPE), was used as the gold standard. This was then compared with the localization results obtained by the 3D-CDD. It was found that the localization coordinates showed a difference of less than 10 mm. |
---|---|
ISSN: | 0309-1902 1464-522X |
DOI: | 10.1080/030919099294258 |