Measured overall perturbation factors at depths greater than dmax for ionization chambers in electron beams
In electron beam dosimetry the perturbation effect in the medium by the ionization chamber cavity is accounted for by introducing a replacement correction factor, Prepl. Another perturbation correction factor, denoted as Pwall, is due to the materials of the walls of the parallel‐plate chamber diffe...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 1999-02, Vol.26 (2), p.208-213 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In electron beam dosimetry the perturbation effect in the medium by the ionization chamber cavity is accounted for by introducing a replacement correction factor, Prepl. Another perturbation correction factor, denoted as Pwall, is due to the materials of the walls of the parallel‐plate chamber differing from the phantom material. Because of the difficulties in separating these two components, we measure the overall perturbation factor, pq=PreplPwall. A distinct advantage of parallel‐plate ionization chambers over cylindrical chambers is that pq has been shown to be close to unity at the standard calibration depth, dmax. However, for many dosimetry applications it is necessary to know the overall perturbation factor at depths greater than dmax. We measured the overall perturbation factor at depths greater than dmax (approximating the 95%, 90% and 50% depth dose) for a Farmer‐type cylindrical ionization chamber and three parallel‐plate ionization chambers. We assumed that pq for the NACP chamber is unity at these measurement depths. The depth dependence for the other chambers was then measured relative to the NACP chamber. The mean energy at depth, Ed, and percentage depth dose gradient ranges studied were 1.9–18.5 MeV and 0 to 4.5%/mm, respectively. For the other two parallel‐plate chambers, we find pq to be unity at depths where the percent depth dose is greater than 90%, but it deviates from unity at deeper depths, where the dose gradients exceed about 2.5%/mm. For the cylindrical chamber, pq values at depths greater than dmax were found to be in good agreement with those in TG 21, where the energy at depth, Ed, is used to evaluate pq. |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1118/1.598506 |