Enhanced dosimetry procedures and assessment for EBT2 radiochromic film

Purpose: Quantitatively determine an optimum image analysis procedure to mitigate inhomogeneities within the EBT2 film and from scanning for accurate absolute dose measurement deposited by an external radiation therapy beam. Multichannel dosimetry procedures were conceived, described, and quantitati...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2012-04, Vol.39 (4), p.2147-2155
Hauptverfasser: Mayer, Rulon R., Ma, Fuhwa, Chen, Yu, Miller, Rachel I., Belard, Arnaud, McDonough, James, O’Connell, John J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose: Quantitatively determine an optimum image analysis procedure to mitigate inhomogeneities within the EBT2 film and from scanning for accurate absolute dose measurement deposited by an external radiation therapy beam. Multichannel dosimetry procedures were conceived, described, and quantitatively tested against single and dual channel dosimetry. Methods: A solid waterTM block was placed on CT imaging and treatment tables in a configuration that avoids bulky compressive devices. CT markers helped register the CT to the treatment plan and the radiation dose distribution from the radiochromic film. The CT images were digitally rotated and resampled to match the spatial resolution of the scanned dosimetric distribution and treatment plan. The ECLIPSE treatment plan planes were digitally translated through digital triangulation of the treatment isocenter to the CT markers in the CT image. A 6 MV photon beam, conforming to the treatment plan, irradiated the EBT2 film sandwiched between solid waterTM slabs. The exposed radiochromic film images were rotated and translated to the CT images using coincident markers in the CT image that are associated with “tattoos” marked on the radiochromic film. The exposed radiochromic film gray-levels from a flatbed scanner in reflection mode were converted to dose using calibration films. The test dose distribution was scanned and averaged six times to reduce temporal noise. This study generated dose distributions using the red channel alone, green channel alone, ratio of the red to blue channel, ratio of the green to blue channel, a hybrid approach combining the green to blue ratio for higher doses (>80 cGy) with the red to blue ratio (
ISSN:0094-2405
2473-4209
DOI:10.1118/1.3694100