Rapid microarray processing using a disposable hybridization chamber with an integrated micropump
We present a disposable microarray hybridization chamber with an integrated micropump to speed up diffusion based reaction kinetics by generating convective flow. The time-to-result for the hybridization reaction was reduced from 60 min (standard protocol) down to 15 min for a commercially available...
Gespeichert in:
Veröffentlicht in: | Lab on a chip 2012-04, Vol.12 (7), p.1384-1388 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present a disposable microarray hybridization chamber with an integrated micropump to speed up diffusion based reaction kinetics by generating convective flow. The time-to-result for the hybridization reaction was reduced from 60 min (standard protocol) down to 15 min for a commercially available microarray. The integrated displacement micropump is pneumatically actuated. It includes two active microvalves and is designed for low-cost, high volume manufacturing. The setup is made out of two microstructured polymer parts realized in polycarbonate (PC) separated by a 25 μm thermoplastic elastomer (TPE) membrane. Pump rate can be controlled between 0.3 μl s(-1) and 5.7 μl s(-1) at actuation frequencies between 0.2 Hz and 8.0 Hz, respectively. |
---|---|
ISSN: | 1473-0197 1473-0189 |
DOI: | 10.1039/c2lc21110f |