Calibration of activity readings from radiopharmaceuticals with impurities
Accurately determining the amount of a radionuclide actually administered to a patient is an important aspect of quality control in nuclear medicine. In this study, in order to accurately assess the “apparent” radioactivities of radionuclides with impurities by a dose calibrator, two methods, (1) cu...
Gespeichert in:
Veröffentlicht in: | Journal of radioanalytical and nuclear chemistry 2011-10, Vol.290 (1), p.153-157 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Accurately determining the amount of a radionuclide actually administered to a patient is an important aspect of quality control in nuclear medicine. In this study, in order to accurately assess the “apparent” radioactivities of radionuclides with impurities by a dose calibrator, two methods, (1) curve fitting, and (2) relative sensitivities calculation, have been employed to determine the parameter,
M
(relative response). From the comparisons between the measurement and re-calculated radioactivity using the
M
values, each method was found to have its own advantage and disadvantage, depending upon the fraction ratio of the impurity as well as the energy level of the impurity’s photons. This study might be helpful for improving the assessment of radiopharmaceuticals’ “true” activity and its “apparent” activity measured by a dose calibrator. |
---|---|