Solid-state photon-counting hybrid detector array for high-resolution multi-energy X-ray imaging
We present in this article the development of a photon-counting, energy-discriminating modular detector based on a pixelated CdZnTe sensor coupled pixel-by-pixel to a novel Digital Pixel Sensor (DPS) readout. The detector is designed for munitions inspection, breast X-ray CT and SPECT/MRI. The curre...
Gespeichert in:
Veröffentlicht in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2011-10, Vol.652 (1), p.470-473 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present in this article the development of a photon-counting, energy-discriminating modular detector based on a pixelated CdZnTe sensor coupled pixel-by-pixel to a novel Digital Pixel Sensor (DPS) readout. The detector is designed for munitions inspection, breast X-ray CT and SPECT/MRI. The current DPS design can also be used to read out other solid-state sensors. The prototype detector is 5.5
mm×5.5
mm in size, and consists of 19×19 pixels on a 250
μm pitch. The DPS is designed in a 0.35
μm process, and every pixel includes a preamplifier, a leakage-current subtraction circuit, an auto-zeroed programmable-gain stage, five comparators, a variable-delay reset circuit and five 16
bit counters. The module is expected to operate at high X-ray fluence exceeding 80
MHz/mm
2, and to improve resolution and contrast in images, while significantly enhancing their signal-to-noise ratio, and assist in identifying material composition via dual-energy imaging. The detector design, fabrication and anticipated performance are discussed. |
---|---|
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/j.nima.2011.01.135 |