Induced radioactive nuclides of 10-MeV radiotherapy accelerators detected by using a portable HP-Ge survey meter

The radioactivation of linear accelerator components for radiation therapy is interest for radiation protection in general, and particularly, when decommissioning these structures. The energy spectra of gamma rays emitted from the heads of two accelerator models, EXL-15SP and Clinac iX, after 10-MeV...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation protection dosimetry 2012-01, Vol.148 (2), p.168-173
Hauptverfasser: Fujibuchi, Toshioh, Obara, Satoshi, Yamaguchi, Ichiro, Oyama, Masaya, Watanabe, Hiroshi, Sakae, Takeji, Katoh, Kazuaki
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The radioactivation of linear accelerator components for radiation therapy is interest for radiation protection in general, and particularly, when decommissioning these structures. The energy spectra of gamma rays emitted from the heads of two accelerator models, EXL-15SP and Clinac iX, after 10-MeV X-ray irradiation, were measured using a high-purity germanium semiconductor survey meter. After spectrum analyses, activities of 24Na, 28Al, 54Mn, 56Mn, 57Ni, 58Co, 60Co, 64Cu, 65Zn, 122Sb, 124Sb, 181W, 187W, 196Au, and 198Au were detected. One centimetre deep dose-equivalent rate of the heads of the linear accelerator was measured using the survey meter. The dose rate decreased to ∼10 % of its initial rate after 1 week. Long-term activations were few, the radioactivity level was low, and a cooling time of several days was effective for reducing dose rate to an acceptable level for decommissioning.
ISSN:0144-8420
1742-3406
DOI:10.1093/rpd/ncr005