Transcatheter Heart Valve With Variable Geometric Configuration: In Vitro Evaluation

Clinically, the current transcatheter aortic valve (TAV) technology has shown a propensity for paravalvular leakage; studies have correlated this flaw to increased calcification at the implantation site and with nonideal geometry of the stented valve. The present study evaluated the hydrodynamics of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Artificial organs 2011-12, Vol.35 (12), p.1151-1159
Hauptverfasser: Young, Ernest, Chen, Ji-Feng, Dong, Owen, Gao, Shengqiang, Massiello, Alex, Fukamachi, Kiyotaka
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Clinically, the current transcatheter aortic valve (TAV) technology has shown a propensity for paravalvular leakage; studies have correlated this flaw to increased calcification at the implantation site and with nonideal geometry of the stented valve. The present study evaluated the hydrodynamics of different geometric configurations, in particular the intravalvular considerations. Three TAV devices were made to create a representative, size 26 mm TAV. Hydrodynamics were assessed using a pulse duplicator. The geometries tested were composed of the nominal, elliptical, triangular, and undersized shapes; along with half‐constriction, a conformation in which only a portion of the stent was constrained. The TAVs were assessed for transvalvular pressure gradient (TVG), effective orifice area (EOA), and regurgitant fraction. The nominal‐sized shape posed a larger TVG (6.2 ± 0.3  mm Hg) than other configurations (P 
ISSN:0160-564X
1525-1594
DOI:10.1111/j.1525-1594.2011.01331.x