FLASH correlation: a new method for 3-D ultrasound tissue motion tracking and blood velocity estimation

We report a new method for tracking tissue motion and blood flow in three dimensions (3-D) using successive volumetric ultrasound scans. The method is based on combining the concepts of feature tracking and 3-D correlation search to achieve a compromise between accuracy and computational efficiency....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control ferroelectrics, and frequency control, 1999, Vol.46 (3), p.728-736
Hauptverfasser: Morsy, A.A., Von Ramm, O.T.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report a new method for tracking tissue motion and blood flow in three dimensions (3-D) using successive volumetric ultrasound scans. The method is based on combining the concepts of feature tracking and 3-D correlation search to achieve a compromise between accuracy and computational efficiency. This paper introduces the new method and the experimental setup used for its evaluation in a tissue-mimicking material. Results are presented demonstrating that the new method has both satisfactory tracking performance and the potential for practical real-time implementation.
ISSN:0885-3010
1525-8955
DOI:10.1109/58.764859