Lower back-up rates improve ventilator triggering during assist-control ventilation: a randomized crossover trial
Objective: The objective of this study is to compare the effects of back-up ventilation rates (BURs) on triggered inflations and patient cardiorespiratory stability during assist-control/volume guarantee ventilation (AC/VG). Study Design: This study is a randomized crossover trial conducted in a neo...
Gespeichert in:
Veröffentlicht in: | Journal of perinatology 2012-02, Vol.32 (2), p.111-116 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective:
The objective of this study is to compare the effects of back-up ventilation rates (BURs) on triggered inflations and patient cardiorespiratory stability during assist-control/volume guarantee ventilation (AC/VG).
Study Design:
This study is a randomized crossover trial conducted in a neonatal unit in an Australian tertiary NICU. In all, 26 stable preterm infants on AC/VG ventilation were studied at BUR settings of 30, 40 and 50 min
−1
. Inflation rate, triggering and cardiorespiratory measures of patient stability were compared during 20 min epochs with 10 min washout periods.
Result:
The 26 infants studied were median (inter-quartile range) gestational age 27 (26, 30) weeks, birth weight 0.84 (0.75, 1.14) kg and FiO
2
0.24 (0.21, 0.31) and age 6 (4, 19) days. At BURs of 30, 40 and 50, the proportions of inflations, which were triggered, were mean (s.d.) 85% (11), 75% (19) and 61% (25);
P |
---|---|
ISSN: | 0743-8346 1476-5543 |
DOI: | 10.1038/jp.2011.73 |