Status of radiation protection in interventional cardiology in four East European countries
Level of staff and patient radiation protection in interventional cardiology in four counties (Bosnia and Herzegovina, Croatia, Montenegro and Serbia) as a part of International Atomic Energy Agency project (RER/9/093) are presented. Patient doses were assessed in terms of air kerma area product (KA...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2011-09, Vol.147 (1-2), p.62-67 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Level of staff and patient radiation protection in interventional cardiology in four counties (Bosnia and Herzegovina, Croatia, Montenegro and Serbia) as a part of International Atomic Energy Agency project (RER/9/093) are presented. Patient doses were assessed in terms of air kerma area product (KAP), peak skin dose (PSD) or air kerma at interventional reference point (K
IRP). Results were available from nine hospitals: 775 patients for KAP, 157 for PSD and 437 for K
IRP. Eight centres reported KAP >100 Gy cm2 and five centres reported values >200 Gy cm2. From patients monitored in terms of PSD, 14 (9 %) had PSD >2 Gy and 6 (3 %) patients from those monitored in terms of K
IRP had value >5 Gy, indicating risk of skin injury. The results indicate need for optimisation and dose monitoring in complex fluoroscopically guided cardiology interventions. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncr268 |