A new method for determining the efficiency of large-area beta sources constructed from anodized aluminum foils

A new method has been developed for determining the efficiency of large-area beta sources in anodized aluminum foils using transmission measurements. The method was applied to the efficiency measurement of a 90Sr– 90Y large-area reference source. Measurement results show that the method can provide...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied radiation and isotopes 2011, Vol.69 (1), p.227-230
Hauptverfasser: Stanga, D., Maringer, F.J., Ionescu, E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new method has been developed for determining the efficiency of large-area beta sources in anodized aluminum foils using transmission measurements. The method was applied to the efficiency measurement of a 90Sr– 90Y large-area reference source. Measurement results show that the method can provide efficiency values for 90Sr– 90Y reference sources with standard uncertainties smaller than 2.9%, which are far below the limit of 10% required by ISO 8769.
ISSN:0969-8043
1872-9800