Comparison of Real-Time In Vivo Spectral and Vector Velocity Estimation

Abstract The purpose of this study is to show whether a newly introduced vector flow method is equal to conventional spectral estimation. Thirty-two common carotid arteries of 16 healthy volunteers were scanned using a BK Medical ProFocus scanner (DK-2730, Herlev, Denmark) and a linear transducer at...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Ultrasound in medicine & biology 2012, Vol.38 (1), p.145-151
Hauptverfasser: Pedersen, Mads Møller, Pihl, Michael Johannes, Haugaard, Per, Hansen, Jens Munk, Hansen, Kristoffer Lindskov, Nielsen, Michael Bachmann, Jensen, Jørgen Arendt
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract The purpose of this study is to show whether a newly introduced vector flow method is equal to conventional spectral estimation. Thirty-two common carotid arteries of 16 healthy volunteers were scanned using a BK Medical ProFocus scanner (DK-2730, Herlev, Denmark) and a linear transducer at 5 MHz. A triplex imaging sequence yields both the conventional velocity spectrum and a two-dimensional vector velocity image. Several clinical parameters were estimated and compared for the two methods: Flow angle, peak systole velocity (PS), end diastole velocity (ED) and resistive index (RI). With a paired t -test, the spectral and vector angles did not differ significantly ( p  = 0.658), whereas PS ( p  = 0.034), ED ( p  = 0.004) and RI ( p < 0.0001) differed significantly. Vector flow can measure the angle for spectral angle correction, thus eliminating the bias from the radiologist performing the angle setting with spectral estimation. The flow angle limitation in velocity estimation is also eliminated, so that flow at any angle can be measured.
ISSN:0301-5629
1879-291X
DOI:10.1016/j.ultrasmedbio.2011.10.003