Use of a remotely steerable “robotic” catheter in a branched endovascular aortic graft

We report the use of a remotely steerable catheter to treat kinked renal bridging stents 8 months after branched endovascular repair of a type III thoracoabdominal aortic aneurysm. Conventional techniques using single, coaxial, and manually steerable sheaths proved too unstable to provide the suppor...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of vascular surgery 2012, Vol.55 (1), p.223-225
Hauptverfasser: Carrell, Tom, MA, MChir, FRCS, Dastur, Neville, FRCS, Salter, Richard, FRCR, Taylor, Peter, MA, MChir, FRCS
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report the use of a remotely steerable catheter to treat kinked renal bridging stents 8 months after branched endovascular repair of a type III thoracoabdominal aortic aneurysm. Conventional techniques using single, coaxial, and manually steerable sheaths proved too unstable to provide the support required to pass a wire against resistance through the kinked stent. A remotely steerable “robotic” catheter provided sufficient precision and stability to cross the kink and reline it with an additional stent, restoring renal perfusion. This technology can help achieve precise and stable introducer sheath position. Further evaluation is necessary to understand the wider applications.