Comparative study of application accuracy of two frameless neuronavigation systems: experimental error assessment quantifying registration methods and clinically influencing factors

This study aimed at comparing the accuracy of two commercial neuronavigation systems. Error assessment and quantification of clinical factors and surface registration, often resulting in decreased accuracy, were intended. Active (Stryker Navigation) and passive (VectorVision Sky, BrainLAB) neuronavi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Neurosurgical review 2011-04, Vol.34 (2), p.217-228
Hauptverfasser: Paraskevopoulos, Dimitrios, Unterberg, Andreas, Metzner, Roland, Dreyhaupt, Jens, Eggers, Georg, Wirtz, Christian Rainer
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study aimed at comparing the accuracy of two commercial neuronavigation systems. Error assessment and quantification of clinical factors and surface registration, often resulting in decreased accuracy, were intended. Active (Stryker Navigation) and passive (VectorVision Sky, BrainLAB) neuronavigation systems were tested with an anthropomorphic phantom with a deformable layer, simulating skin and soft tissue. True coordinates measured by computer numerical control were compared with coordinates on image data and during navigation, to calculate software and system accuracy respectively. Comparison of image and navigation coordinates was used to evaluate navigation accuracy. Both systems achieved an overall accuracy of
ISSN:0344-5607
1437-2320
DOI:10.1007/s10143-010-0302-5