Improved Reflexive Testing Algorithm for Hepatitis C Infection Using Signal-to-Cutoff Ratios of a Hepatitis C Virus Antibody Assay

Chemiluminescence immunoassay (CIA) is used to detect hepatitis C virus (HCV) antibody status on the basis of signal-to-cutoff (S/Co) ratios. Positive results of antibody to HCV (anti-HCV) are followed by either recombinant immunoblot assay (RIBA) to confirm anti-HCV positivity or reverse transcript...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical chemistry (Baltimore, Md.) Md.), 2011-07, Vol.57 (7), p.1050-1056
Hauptverfasser: LAI, Keane K. Y, MING JIN, SHAN YUAN, LARSON, Meaghan F, DOMINITZ, Jason A, BANKSON, Daniel D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Chemiluminescence immunoassay (CIA) is used to detect hepatitis C virus (HCV) antibody status on the basis of signal-to-cutoff (S/Co) ratios. Positive results of antibody to HCV (anti-HCV) are followed by either recombinant immunoblot assay (RIBA) to confirm anti-HCV positivity or reverse transcription (RT)-PCR to detect viremia. We hypothesized that by analyzing S/Co ratios, we could determine a strategy to reduce unnecessary supplementary testing in our population. CIA was performed to screen for anti-HCV, and positive results were followed up with RT-PCR testing. Negative RT-PCR results were followed up with RIBA, whereas positive RT-PCR results were assumed to be RIBA positive. ROC curves were analyzed to determine the optimal S/Co ratios to predict HCV infection. We determined the S/Co ratios on 34 243 veteran patient samples. We found that with the CIA method 9.0% of patients had positive test results for anti-HCV. An S/Co ratio
ISSN:0009-9147
1530-8561
DOI:10.1373/clinchem.2010.158691