Small Mass-Change Detectable Quartz Crystal Microbalance and Its Application to Enzymatic One-Base Elongation on DNA
A highly sensitive 27 MHz quartz crystal microbalance instrument with an automatic flow injection system was developed to obtain realistic minimal frequency noise (±0.05 Hz) and to obtain a stable signal baseline (±1 Hz/h) by controlling the temperature of each part in the quartz crystal microbalanc...
Gespeichert in:
Veröffentlicht in: | Analytical chemistry (Washington) 2011-11, Vol.83 (22), p.8741-8747 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A highly sensitive 27 MHz quartz crystal microbalance instrument with an automatic flow injection system was developed to obtain realistic minimal frequency noise (±0.05 Hz) and to obtain a stable signal baseline (±1 Hz/h) by controlling the temperature of each part in the quartz crystal microbalance (QCM) system using three Peltier devices with a resolution of ±0.001 °C and by optimizing the flow system to prevent fluctuation of the internal pressure of the QCM. The improved QCM with an automatic flow injection system enabled detection of small mass changes such as binding of biotin to a streptavidin-immobilized QCM with a high signal-to-noise ratio. We also applied this device to enzyme reactions of one-base elongation by DNA polymerase (Klenow fragment, KF). We immobilized dsDNAs including the protruding end of dA, dG, dT, or dC on the QCM electrode and ran complementary dNTP monomers with KF into the QCM flow cell. We could directly detect the enzymatic one-base elongation of DNA as a small mass increase, and we found the difference in the reaction rate for each monomer. |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/ac202224d |