Analysis of cyclosporin A in dried blood spots using liquid chromatography tandem mass spectrometry
Dried blood spot sampling is a promising and patient friendly alternative for venous sampling. A liquid chromatography tandem mass spectrometer assay was developed for analyzing cyclosporin A in dried blood spots. Linearity ranged from 25 to 1440μg/L. Within and between run accuracy and precision we...
Gespeichert in:
Veröffentlicht in: | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2009-05, Vol.877 (14-15), p.1595-1598 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Dried blood spot sampling is a promising and patient friendly alternative for venous sampling. A liquid chromatography tandem mass spectrometer assay was developed for analyzing cyclosporin A in dried blood spots. Linearity ranged from 25 to 1440μg/L. Within and between run accuracy and precision were within limits. The developed assay has a negligible matrix-effect and a recovery of 97%. The dried blood spots were stable during a period of at least 17 days in the refrigerator. The developed assay is suitable for analyzing cyclosporin A in dried blood spots. |
---|---|
ISSN: | 1570-0232 1873-376X |
DOI: | 10.1016/j.jchromb.2009.03.024 |