Real-time measurement of peroxyacetyl nitrate using selected ion flow tube mass spectrometry

The on‐line detection of gaseous peroxyacetyl nitrate (PAN) using selected ion flow tube mass spectrometry (SIFT‐MS) has been investigated using a synthetic sample of PAN in air at a humidity of ∼30%. Using the H3O+ reagent ion, signals due to PAN at m/z 122, 77 and 95 have been identified. These co...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Rapid communications in mass spectrometry 2010-02, Vol.24 (3), p.343-348
Hauptverfasser: Hastie, Donald R., Gray, John, Langford, Vaughan S., Maclagan, Robert G. A. R., Milligan, Daniel B., McEwan, Murray J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The on‐line detection of gaseous peroxyacetyl nitrate (PAN) using selected ion flow tube mass spectrometry (SIFT‐MS) has been investigated using a synthetic sample of PAN in air at a humidity of ∼30%. Using the H3O+ reagent ion, signals due to PAN at m/z 122, 77 and 95 have been identified. These correspond to protonated PAN, protonated peractetic acid and its water cluster, respectively. These products and their energetics have been probed through quantum mechanical calculations. The rate coefficient of H3O+ has been estimated to be 4.5 × 10−9 cm3 s−1, leading to a PAN sensitivity of 138 cps/ppbv. This gives a limit of detection of 20 pptv in 10 s using the [M+H]+ ion of PAN at m/z 122. Copyright © 2010 John Wiley & Sons, Ltd.
ISSN:0951-4198
1097-0231
1097-0231
DOI:10.1002/rcm.4400