Three dimension optical imaging in a high throughput layered expression system
Layered peptide array (LPA) system enables multiplex screening of biomarkers [1-3]. One of the main problems of the LPA system is the screening of the layered-membranes stack. Nowadays, each membrane is imaged separately using conventional fluorescent imaging. This process is time consuming and requ...
Gespeichert in:
Veröffentlicht in: | Optics express 2011-10, Vol.19 (21), p.19822-19835 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Layered peptide array (LPA) system enables multiplex screening of biomarkers [1-3]. One of the main problems of the LPA system is the screening of the layered-membranes stack. Nowadays, each membrane is imaged separately using conventional fluorescent imaging. This process is time consuming and requires extensive manual interaction. This paper describes a general solution for optical imaging of a layered grid medium using photogrammetric methods. The suggested method enables visualization of the LPA membranes stack by using only two images of the stack. This study is a proof of concept of the suggested solution using MATLAB simulation and phantom experiments. |
---|---|
ISSN: | 1094-4087 1094-4087 |
DOI: | 10.1364/OE.19.019822 |