Noninvasive estimation of dynamic pressures in vitro and in vivo using the subharmonic response from microbubbles

The purpose of this study was to develop and validate a noninvasive pressure estimation technique based on subharmonic emissions from a commercially available ultrasound contrast agent and scanner, unlike other studies that have either adopted a single-element transducer approach and/ or use of in-h...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control ferroelectrics, and frequency control, 2011-10, Vol.58 (10), p.2056-2066
Hauptverfasser: Dave, J. K., Halldorsdottir, V. G., Eisenbrey, J. R., Ji-Bin Liu, McDonald, M. E., Dickie, K., Leung, C., Forsberg, F.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to develop and validate a noninvasive pressure estimation technique based on subharmonic emissions from a commercially available ultrasound contrast agent and scanner, unlike other studies that have either adopted a single-element transducer approach and/ or use of in-house contrast agents. Ambient pressures were varied in a closed-loop flow system between 0 and 120 mmHg and were recorded by a solid-state pressure catheter as the reference standard. Simultaneously, the ultrasound scanner was operated in pulse inversion mode transmitting at 2.5 MHz, and the unprocessed RF data were captured at different incident acoustic pressures (from 76 to 897 kPa). The subharmonic data for each pulse were extracted using band-pass filtering with averaging, and subsequently processed to eliminate noise. The incident acoustic pressure most sensitive to ambient pressure fluctuations was determined, and then the ambient pressure was tracked over 20 s. In vivo validation of this technique was performed in the left ventricle (LV) of 2 canines. In vitro, the subharmonic signal could track ambient pressure values with r 2 = 0.922 (p ; 0.790 (p
ISSN:0885-3010
1525-8955
DOI:10.1109/TUFFC.2011.2056