A three-dimensional neural recording microsystem with implantable data compression circuitry

A 256-site, fully implantable, 3-D neural recording microsystem has been developed. The microsystem incorporates four active neural probes with integrated circuitry for site selection, amplification, and multiplexing. The probes drive an embedded data-compression ASIC that successfully detects neura...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE journal of solid-state circuits 2005-12, Vol.40 (12), p.2796-2804
Hauptverfasser: Olsson, R.H., Wise, K.D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A 256-site, fully implantable, 3-D neural recording microsystem has been developed. The microsystem incorporates four active neural probes with integrated circuitry for site selection, amplification, and multiplexing. The probes drive an embedded data-compression ASIC that successfully detects neural spikes in the presence of neural and circuit noise. The spike detection ASIC achieves a factor of 12 bandwidth reduction while preserving the key features of the action potential waveshape necessary for spike discrimination. This work extends the total number of neural channels that can be recorded across a transcutaneous inductively coupled wireless link from 25 to 312. When a spike is detected, this ASIC serially shifts the 5-bit amplitude and 5-bit address of the spike off of the chip over a single 2.5 Mb/s wired or wireless line. The spike detection ASIC occupies 6 mm/sup 2/ in 0.5 /spl mu/m features and consumes 2.6 mW while the entire microsystem consumes 5.4 mW of power from a 3-V supply.
ISSN:0018-9200
1558-173X