Interlaboratory comparison at 10 V DC
An interlaboratory comparison at 10 V DC has been made among 16 national, industrial, and military standards laboratories in North America. The comparison used a set of four traveling Zener dc reference standards. A pivot laboratory made measurements at the beginning, at the end, and at four other t...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on instrumentation and measurement 2005-01, Vol.54 (1), p.215-221 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An interlaboratory comparison at 10 V DC has been made among 16 national, industrial, and military standards laboratories in North America. The comparison used a set of four traveling Zener dc reference standards. A pivot laboratory made measurements at the beginning, at the end, and at four other times during the comparison. The measured differences and their uncertainties are reported and used to demonstrate equivalence between each participant and the pivot, and between each participant and the National Institute of Standards and Technology (NIST). All of the differences fall within two parts in 10/sup 8/. |
---|---|
ISSN: | 0018-9456 1557-9662 |