Transverse relaxometry with stimulated echo compensation

Presented is a fitting model for transverse relaxometry data acquired with the multiple‐refocused spin‐echo sequence. The proposed model, requiring no additional data input or pulse sequence modifications, compensates for imperfections in the transmit field and radiofrequency (RF) profiles. Exploiti...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 2010-10, Vol.64 (4), p.1005-1014
Hauptverfasser: Lebel, R. Marc, Wilman, Alan H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Presented is a fitting model for transverse relaxometry data acquired with the multiple‐refocused spin‐echo sequence. The proposed model, requiring no additional data input or pulse sequence modifications, compensates for imperfections in the transmit field and radiofrequency (RF) profiles. Exploiting oscillatory echo behavior to estimate alternate coherence pathways, the model compensates for prolonged signal decay from stimulated echo pathways yielding precise monoexponential T2 quantification. Verified numerically and experimentally at 4.7 T in phantoms and the human brain, over 95% accuracy is readily attainable in realistic imaging situations without sacrificing multislice capabilities or requiring composite or adiabatic RF pulses. The proposed model allows T2 quantitation in heterogeneous transmit fields and permits thin refocusing widths for efficient multislice imaging. Magn Reson Med, 2010. © 2010 Wiley‐Liss, Inc.
ISSN:0740-3194
1522-2594
1522-2594
DOI:10.1002/mrm.22487