A technique for rapid single-echo spin-echo T2 mapping

A rapid technique for mapping of T2 relaxation times is presented. The method is based on the conventional single‐echo spin echo approach but uses a much shorter pulse repetition time to accelerate data acquisition. The premise of the new method is the use of a constant difference between the echo t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 2010-08, Vol.64 (2), p.536-545
Hauptverfasser: Sussman, Marshall S., Vidarsson, Logi, Pauly, John M., Cheng, Hai-Ling Margaret
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A rapid technique for mapping of T2 relaxation times is presented. The method is based on the conventional single‐echo spin echo approach but uses a much shorter pulse repetition time to accelerate data acquisition. The premise of the new method is the use of a constant difference between the echo time and pulse repetition time, which removes the conventional and restrictive requirement of pulse repetition time ≫ T1. Theoretical and simulation investigations were performed to evaluate the criteria for accurate T2 measurements. Measured T2s were shown to be within 1% error as long as the key criterion of pulse repetition time/T2 ≥3 is met. Strictly, a second condition of echo time/T1 ≪ 1 is also required. However, violations of this condition were found to have minimal impact in most clinical scenarios. Validation was conducted in phantoms and in vivo T2 mapping of healthy cartilage and brain. The proposed method offers all the advantages of single‐echo spin echo imaging (e.g., immunity to stimulated echo effects, robustness to static field inhomogeneity, flexibility in the number and choice of echo times) in a considerably reduced amount of time and is readily implemented on any clinical scanner. Magn Reson Med, 2010. © 2010 Wiley‐Liss, Inc.
ISSN:0740-3194
1522-2594
1522-2594
DOI:10.1002/mrm.22454