Replacement-free mass-amplified sandwich assay with 180-MHz electrodeless quartz-crystal microbalance biosensor

This study describes a sensitivity-amplified detection method in a replacement-free electrodeless quartz-crystal microbalance (QCM) biosensor. A sandwich assay is proposed for detecting C-reactive proteins (CRP), where the biotinated second anti-CRP antibody is weighted by streptavidin for sensitivi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biosensors & bioelectronics 2011-08, Vol.26 (12), p.4819-4822
Hauptverfasser: Ogi, Hirotsugu, Yanagida, Taiji, Hirao, Masahiko, Nishiyama, Masayoshi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study describes a sensitivity-amplified detection method in a replacement-free electrodeless quartz-crystal microbalance (QCM) biosensor. A sandwich assay is proposed for detecting C-reactive proteins (CRP), where the biotinated second anti-CRP antibody is weighted by streptavidin for sensitivity amplification. Because the first CRP antibody was immobilized nonspecifically on naked quartz surfaces, the sandwich assay was repeated using the same sensor chip, making possible the replacement-free assay. The mass-amplified sandwich assay detected a CRP solution of 0.1 ng/ml. A methodology for determining the molecular mass of the injected protein is also proposed.
ISSN:0956-5663
1873-4235
DOI:10.1016/j.bios.2011.05.035