Experimental Comparative Study of Efficiency of Three Methods for Measuring Resolution of Digital Radiographic Systems

An experimental comparative study of the efficiency of three methods for measurement of the resolution of digital radiographic systems is described. An algorithm for subpixel angular projection was constructed. This algorithm allows the resolution of X-ray camera and test object to be measured using...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical engineering 2011-07, Vol.45 (2), p.39-42
Hauptverfasser: Gurzhiev, A. S., Gurzhiev, S. N., Miroshnikov, E. M., Novikov, V. P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An experimental comparative study of the efficiency of three methods for measurement of the resolution of digital radiographic systems is described. An algorithm for subpixel angular projection was constructed. This algorithm allows the resolution of X-ray camera and test object to be measured using a narrow slit and sharp edge. Test results obtained using various methods are compared.
ISSN:0006-3398
1573-8256