Evaluation of absolute peptide quantitation strategies using selected reaction monitoring

The use of internal peptide standards in selected reaction monitoring experiments enables absolute quantitation. Here, we describe three approaches addressing calibration of peptide concentrations in complex matrices and assess their performance in terms of trueness and precision. The simplest appro...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Proteomics (Weinheim) 2011-03, Vol.11 (6), p.1148-1152
Hauptverfasser: Campbell, James, Rezai, Taha, Prakash, Amol, Krastins, Bryan, Dayon, Loïc, Ward, Malcolm, Robinson, Sarah, Lopez, Mary
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The use of internal peptide standards in selected reaction monitoring experiments enables absolute quantitation. Here, we describe three approaches addressing calibration of peptide concentrations in complex matrices and assess their performance in terms of trueness and precision. The simplest approach described is single reference point quantitation where a heavy peptide is spiked into test samples and the endogenous analyte quantified relative to the heavy peptide internal standard. We refer to the second approach as normal curve quantitation. Here, a constant amount of heavy peptide and a varying amount of light peptide are spiked into matrix to construct a calibration curve. This accounts for matrix effects but due to the presence of endogenous analyte, it is usually not possible to determine the lower LOQ. We refer to the third method as reverse curve quantitation. Here, a constant amount of light peptide and a varying amount of heavy peptide are spiked into matrix to construct a calibration curve. Because there is no contribution to the heavy peptide signal from endogenous analyte, it is possible to measure the equivalent of a blank sample and determine LOQ. These approaches are applied to human plasma samples and used to assay peptides of a set of apolipoproteins.
ISSN:1615-9853
1615-9861
1615-9861
DOI:10.1002/pmic.201000511