Use of Near-Infrared Spectroscopy for Estimation of Renal Oxygenation in Children With Heart Disease
We evaluated whether near-infrared spectroscopy (NIRS) measurement from the flank correlates with renal vein saturation in children undergoing cardiac catheterization. Thirty-seven patients 10 kg, there was no correlation between rSO 2 and renal vein saturation ( r = 0.158, p = 0.57) or IVC satur...
Gespeichert in:
Veröffentlicht in: | Pediatric cardiology 2011-08, Vol.32 (6), p.748-753 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We evaluated whether near-infrared spectroscopy (NIRS) measurement from the flank correlates with renal vein saturation in children undergoing cardiac catheterization. Thirty-seven patients 10 kg, there was no correlation between rSO
2
and renal vein saturation (
r
= 0.158,
p
= 0.57) or IVC saturation (
r
= –0.107,
p
= 0.67). Regional tissue oxygenation as measured by flank NIRS correlates well with both renal vein and IVC oxygen saturations in children weighing |
---|---|
ISSN: | 0172-0643 1432-1971 |
DOI: | 10.1007/s00246-011-9960-5 |