Gamma camera intrinsic uniformity in an unstable power supply environment

The main objective of this work was to show that a gamma camera in a developing country could perform efficiently despite electricity outages using intrinsic flood uniformity tests as an index of performance. A total of 143 intrinsic uniformity test results for a new gamma camera in use in an enviro...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Hellenic journal of nuclear medicine 2011-05, Vol.14 (2), p.146-148
Hauptverfasser: Ejeh, John E, Adedapo, Kayode S, Akinlade, Bidemi I, Osifo, Bola O A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The main objective of this work was to show that a gamma camera in a developing country could perform efficiently despite electricity outages using intrinsic flood uniformity tests as an index of performance. A total of 143 intrinsic uniformity test results for a new gamma camera in use in an environment with unstable power supply are presented. The integral uniformity for the central field of view (CFOV) was found to be between 3.43% and 1.49% (3.29% for acceptance test) while the integral uniformity for the useful field of view (UFOV) was between 4.51% and 1.9% (5.21% for acceptance test). The differential uniformity for the CFOV was between 1.99% and 1.04% (2.25% for acceptance test) while that of the UFOV was between 2.84% and 1.23% (2.63% for acceptance test). In conclusion, these results show that the uniformity of the gamma camera under this condition is within an acceptable range for both planar and SPET imaging.
ISSN:1790-5427