Image Quality in Reduced-Dose Coronary CT Angiography

Rationale and Objectives Concerns for patient’s risk of radiation-induced cancer have increased demand for reduced-dose coronary computed tomography angiography (CCTA). Previous comparisons of full and reduced-dose CCTA were not conclusive, because results were compared in different groups of patien...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Academic radiology 2011-08, Vol.18 (8), p.984-990
Hauptverfasser: Gagarina, Nina V., MD, PhD, Irwan, Roy, MSc, PhD, Gordina, Galina, MD, Fominykh, Ekaterina, MD, PhD, Sijens, Paul E., PhD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Rationale and Objectives Concerns for patient’s risk of radiation-induced cancer have increased demand for reduced-dose coronary computed tomography angiography (CCTA). Previous comparisons of full and reduced-dose CCTA were not conclusive, because results were compared in different groups of patients. Presented here are results in patients examined by a widely used full dose CCTA protocol and a new low-dose alternative. Materials and Methods Standard full-dose and low-dose CCTA with tube voltages of 120/100 kV were applied on 70 patients with intermediate probability of coronary artery disease (CAD). Both protocols used prospective electrocardiogram-gated acquisition on a 320-detector row CT scanner, whereas at low-dose CCTA the phase window was increased from 10% to 75% of R-R interval. Results Despite a mean dose reduction of 80%, from 4.9 ± 0.98 to 0.98 ± 0.24 mSv, visual image quality was not significantly affected at the low-dose protocol. Contrast level, image noise, and CNR for both protocols were similar in the majority of coronary segments. CNR for standard and low-dose protocol were 23.7 ± 17.1 and 23.2 ± 26.8, P = NS. Correlation between visual image quality and heart rate variability was strong at low dose: r = −0.58, P = .01, and absent at full dose: r = −0.07, P = .77. Conclusion Image quality of blood vasculature is generally not affected by 80% CCTA dose reduction applied to standard prospective electrocardiogram-gated acquisition. The performance at the low-dose protocol owes to the increased phase window, enhancing image quality at the cost of sensitivity to heart rate variability as compared with standard CCTA.
ISSN:1076-6332
1878-4046
DOI:10.1016/j.acra.2011.03.009