Radioactive uranium measurement in vivo using a handheld interfaced analyzer
A trace uranium (U) detection method was developed with a handheld voltammetric analyzer that was the size of a mobile phone, with working sensors made of simple graphite pencil electrode (PE). The optimum stripping voltammetric conditions were sought, and the following results were obtained: 0.0 to...
Gespeichert in:
Veröffentlicht in: | Environmental toxicology and chemistry 2010-05, Vol.29 (5), p.1025-1030 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A trace uranium (U) detection method was developed with a handheld voltammetric analyzer that was the size of a mobile phone, with working sensors made of simple graphite pencil electrode (PE). The optimum stripping voltammetric conditions were sought, and the following results were obtained: 0.0 to 0.08 ng/L working ranges and a statistically relative standard deviation of 1.78% (RSD; n = 15) at a 10.0 µg/L U spike. The experiment accumulation time used was only 150 s. Under this condition, the diagnostic detection limit approached 0.007 ng/L. The method was applied to soil of a natural rock in a radioactive mineralogy site. Earthworms that resided at this site were assayed. The method was found to be applicable in biological diagnosis or in real‐time in vivo survey. Environ. Toxicol. Chem. 2010;29:1025–1030. © 2010 SETAC |
---|---|
ISSN: | 0730-7268 1552-8618 1552-8618 |
DOI: | 10.1002/etc.131 |