To correct measurement errors in spectral emissivity measurement based on FT-IR in the vacuum ambient, the emissivity calibration model has been constructed, including radiation thermal losses, one dimension heat transfer and radiation thermometry. The emissivity error has been quantitatively analyz...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Hong wai yu hao mi bo xue bao 2011-02, Vol.29 (5), p.367-371
Hauptverfasser: WANG, Zong-Wei, DAI, Jing-Min, HE, Xiao-Wa, YANG, Chun-Ling, PAN, Wei-Dong
Format: Artikel
Sprache:chi ; eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To correct measurement errors in spectral emissivity measurement based on FT-IR in the vacuum ambient, the emissivity calibration model has been constructed, including radiation thermal losses, one dimension heat transfer and radiation thermometry. The emissivity error has been quantitatively analyzed. The results show that the higher the sample temperature is, the lower is the value of the emissivity. This result is induced by thermal losses, material thermal conductivity and sample thickness. To reduce the error induced by these factors, the system should be calibrated. The calibrated spectral emissivity of high-purity graphite sample at 2000 degree C agrees very well with the results reported by G. Neuer and G. Jaroma-Weiland.
ISSN:1001-9014
DOI:10.3724/SP.J.1010.2010.00367