Comparison of the Inverness Medical Acceava Strep A Test With the Genzyme OSOM and Quidel QuickVue Strep A Tests
Previous studies of the accuracy of rapid in-office tests for group A Streptococcus had disparate results, ranging from a sensitivity of 70% to more than 90%. The sensitivity and specificity of 3 commercially available Strep A tests were determined in 2 private pediatric office settings. Acceava Str...
Gespeichert in:
Veröffentlicht in: | Clinical pediatrics 2011-04, Vol.50 (4), p.294-296 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Previous studies of the accuracy of rapid in-office tests for group A Streptococcus had disparate results, ranging from a sensitivity of 70% to more than 90%. The sensitivity and specificity of 3 commercially available Strep A tests were determined in 2 private pediatric office settings. Acceava Strep A, Genzyme OSOM Strep A, and the Quidel QuickVue Strep A tests were the representative rapid tests for detection of Streptococcus pyogenes. Overnight culture on standard 5% sheep blood agar was the reference standard for this study. All 3 Clinical Laboratory Improvement Amendments—waived tests had sensitivities and specificities that exceeded 95%. |
---|---|
ISSN: | 0009-9228 1938-2707 |
DOI: | 10.1177/0009922810385675 |