Optimal configuration of the spiral linking technique for tendon repair. Biomechanical evaluation

A previous study described a new spiral linking technique for tendon repairs and demonstrated that it was strong enough to be used in clinical practice as an alternative to the Pulvertaft tendon weave repair. However the repairs were less stiff, needed slightly more tendon length for the same repair...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of hand surgery, European volume European volume, 2011-03, Vol.36 (3), p.210-214
Hauptverfasser: Kulikov, Y. I., Vinogradova, Y., Miles, A. W., Giddins, G.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A previous study described a new spiral linking technique for tendon repairs and demonstrated that it was strong enough to be used in clinical practice as an alternative to the Pulvertaft tendon weave repair. However the repairs were less stiff, needed slightly more tendon length for the same repair and were a little bulkier. In this study two variables have been changed with a view to improving the spiral technique. At first the number of spirals was reduced consecutively, keeping the same number of standard mattress sutures. Once the optimal number of spirals had been identified, repairs with different numbers of sutures were tested using an alternative cross-stitch technique. The spiral repair technique using two spirals linked with six sutures was at least as strong and stiff as a four-weave Pulvertaft technique and was also easier to do.
ISSN:1753-1934
2043-6289
DOI:10.1177/1753193410390759