Gradient RP-HPLC method for the determination of potential impurities in atazanavir sulfate
This paper proposes a simple and selective RP-HPLC method for the determination of process impurities and degradation products (degradants) of atazanavir sulfate (ATV) drug substance. Chromatographic separation was achieved on Ascentis ® Express C8, (150 mm × 4.6 mm, 2.7 μm) column thermostated at 3...
Gespeichert in:
Veröffentlicht in: | Journal of pharmaceutical and biomedical analysis 2011-04, Vol.55 (1), p.31-47 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper proposes a simple and selective RP-HPLC method for the determination of process impurities and degradation products (degradants) of atazanavir sulfate (ATV) drug substance. Chromatographic separation was achieved on Ascentis
® Express C8, (150
mm
×
4.6
mm, 2.7
μm) column thermostated at 30
°C under gradient elution by a binary mixture of potassium dihydrogen phosphate (pH 3.5, 0.02
M) and ACN at a flow rate of 1.0
ml/min. A photodiode array (PDA) detector set at 250
nm was used for detection. Stress testing (forced degradation) of ATV was carried out under acidic, alkaline, oxidative, photolytic, thermal and humidity conditions. In presence of alkali, ATV transformed into cyclized products and the order of degradation reaction is determined by the method of initial rates. The unknown process impurities and alkaline degradants are isolated by preparative LC and characterized by ESI-MS/MS,
1H NMR, and FT-IR spectral data. The developed method is validated with respect to sensitivity (lod and loq), linearity, precision, accuracy and robustness and can be implemented for routine quality control analysis and stability testing of ATV. |
---|---|
ISSN: | 0731-7085 1873-264X |
DOI: | 10.1016/j.jpba.2011.01.002 |