Advantages and Limitations of Small Gauge Vitrectomy

Abstract Small gauge vitrectomy utilizing 23- and 25-gauge instrumentation has definite advantages, but also limitations, due to the physics of smaller instruments and sutureless surgery. Higher infusion and aspiration pressures are needed to remove the vitreous using 23- and 25-gauge probes. The ad...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Survey of ophthalmology 2011-03, Vol.56 (2), p.162-172
1. Verfasser: Thompson, John T., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Small gauge vitrectomy utilizing 23- and 25-gauge instrumentation has definite advantages, but also limitations, due to the physics of smaller instruments and sutureless surgery. Higher infusion and aspiration pressures are needed to remove the vitreous using 23- and 25-gauge probes. The advantages include decreased surgical times, less tissue manipulation, reduced inflammation and pain postoperatively with more rapid visual recovery. A disadvantage is greater instrument flexion than 20-gauge probes, making small gauge vitrectomy more appropriate for indications such as vitreous opacities, epiretinal membranes, macular holes, and simple retinal detachments. There are also some increased complications related to small gauge vitrectomy, including dislocation of cannulas intraoperatively, early postoperative hypotony, choroidal detachment, and possibly an increased risk of infectious endophthalmitis.
ISSN:0039-6257
1879-3304
DOI:10.1016/j.survophthal.2010.08.003