Efficacy and safety of 5-mm-diameter bipolar and ultrasonic shears for cutting carotid arteries of the hybrid pig

Background Experimental data about the efficacy and safety of sealing devices are rare. Therefore, this study investigated these parameters for three commercially available energy-based vascular sealing and cutting systems. Methods In male hybrid pigs, 487 carotid artery segments were sealed and cut...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Surgical endoscopy 2011-02, Vol.25 (2), p.577-585
Hauptverfasser: Mantke, René, Halangk, W., Habermann, A., Peters, B., Konrad, S., Guenther, M., Lippert, H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background Experimental data about the efficacy and safety of sealing devices are rare. Therefore, this study investigated these parameters for three commercially available energy-based vascular sealing and cutting systems. Methods In male hybrid pigs, 487 carotid artery segments were sealed and cut using the harmonic scalpel or several bipolar sealing devices. The sealing failure rate, burst pressure, process time, and extent of lateral thermal damage were analyzed. Results A regular sealing and cutting process in more than 90% of the carotid arteries was found using the following instruments: LS1520, ACE (level 1), ACE (level 3), CS14C (level 1), WAVE (level 1), and WAVE (level 5). The largest failure rate was found for the CS14C device (level 5: initial sealing failure, 21.5%). The maximal mean burst pressure (1727 ± 453 mmHg) was reached using the ACE device (level 1). Significant differences were found in the size of the lateral thermal damage, which a ranged from 2.5 mm (LS1520) to 1.51 mm (CS14C, level 1). The process time ranged widely from 6.8 s (ACE, level 5) to 31.83 s (WAVE, level 1). Conclusion The current study demonstrated that all the tested devices are efficacious and safe in sealing and cutting arteries up to 5 mm in diameter. All the devices showed supraphysiologic mean burst pressures. Differences in failure rate, thermal damage, and process time lead to an advised application of the different systems.
ISSN:0930-2794
1432-2218
DOI:10.1007/s00464-010-1224-6