Multicenter evaluation of a new automated electrochemiluminescence immunoassay for the quantification of testosterone compared to liquid chromatography tandem mass spectrometry
A new electrochemiluminescence testosterone immunoassay (Elecsys® Testo II) was evaluated. Within-run precision, total precision, and method comparisons to the former immunoassay and LC-MS/MS were investigated. Reference intervals were established for adults and children. Imprecision ranged from 0.9...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 2011-02, Vol.44 (2), p.264-267 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new electrochemiluminescence testosterone immunoassay (Elecsys® Testo II) was evaluated.
Within-run precision, total precision, and method comparisons to the former immunoassay and LC-MS/MS were investigated. Reference intervals were established for adults and children.
Imprecision ranged from 0.9% to 8.5%. Method comparisons revealed slopes from 0.68 to 1.32 and coefficients of correlation from 0.893 to 0.991.
Improved agreement with LC-MS/MS was demonstrated, especially for female patients. The assay may be a suitable alternative for laboratories with no access to MS techniques. |
---|---|