Measurement of B-type natriuretic peptide by two assays utilizing antibodies with different epitope specificity
To compare plasma BNP values determined by a conventional-type and a novel-type of BNP assays. Plasma samples (n = 94) from HF patients were analyzed by the novel-type “Single Epitope Sandwich” (SES assay prototype) and the conventional-type Siemens ADVIA Centaur BNP assays. Both assays were calibra...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 2011-02, Vol.44 (2), p.257-259 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To compare plasma BNP values determined by a conventional-type and a novel-type of BNP assays.
Plasma samples (n
=
94) from HF patients were analyzed by the novel-type “Single Epitope Sandwich” (SES assay prototype) and the conventional-type Siemens ADVIA Centaur BNP assays. Both assays were calibrated using recombinant proBNP (expressed in
E. coli).
The SES assay measured 1.2- to 7.2-fold (2.1
±
0.9; mean, SD) greater BNP concentrations compared to the Siemens assay. A subset of six samples (6.4%) demonstrated the largest (3- to 7.2-fold higher) difference.
The SES prototype assay appears to more accurately measure the absolute concentrations of BNP immunoreactive forms. |
---|---|
ISSN: | 0009-9120 1873-2933 |
DOI: | 10.1016/j.clinbiochem.2010.09.030 |